Sensor Specifications

Internal and Fixed External Thermistor Sensor Accuracy

RangeAccuracy
-40 to 0 °C ± 0.3 °C
0 to 70 °C ± 0.2 °C
70 to 85 °C ± 0.3 °C
85 to 100 °C ± 0.5 °C

Fixed External Humidity Sensor Accuracy

10-90 % ± 2% RH, Compensated for -5 + 60 °C ambient temperature

Remote Thermistor Sensors' Ranges and Resolutions

Two different external temperature sensors available

TypeRangesResolutionAccuracy
T1 - 40 to 85 °C < 0.1 °C As for built in sensor
     
 

Remote Pt100 Sensor

Four different temperature ranges available.

TypeRangesResolutionAccuracy
A -200 + 120 °C 0.2 °C -200 - 100 °C ± 2.0 °C
-100 - 40 °C ± 1.0 °C
-40 + 120 °C ± 0.5 °C
B -40 + 120 °C 0.1 °C -40 + 120 °C ± 0.3 °C
C -40 + 280 °C 0.2 °C -40 + 100 °C ± 0.4 °C
+100 + 200 °C ± 0.6 °C
+200 + 280 °C ± 1.0 °C

4-20 milli-Amperes input

Accuracy: ± 1% of range

Connection through supplied 2 meter cable with connector for logger.

Time Accuracy

Synchronisation up to: ± 1 second; ± 1 minute per month.